Terminal filtration for “CLEANROOM” application.

Application areas

  • High air flow
  • Very low pressure drop
  • Individual leakage test on 0.3 µm at 0.45 m/s according to DIN 24.184 (from H13)
  • Individual MPPS efficiency test on 0.12 µm at 0.45 m/s (scanning test) according to EN1822 standard
  • Individual test certificate (scanning test) provided with the filter
  • Double labeling for traceability
  • Tested for food contact according to CE 1935/2004
  • Certified inert to microbial growth (ISO 846-VDI 6022)
Certificates

General Characteristics

EN1822 filtration class: E10-U15
Media: mini-pleated glass microfiber paper
Frame: anodized extruded aluminum
Lut: polyurethane
Separators: continuous thermoplastic wires
Joints: monobloc half-round polyurethane
Grilles: painted metal (upstream and downstream)
Final pressure drop: 600 Pa
Maximum continuous service temperature: 80°C
Relative humidity: 70-80%
Contact our expert
  • THE-LA 150 MM

    Illustration variation
  • E10

    • THE-LA100A

    • Base A (mm)

      305

    • Height B (mm)

      305

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      4,5

    • Airflow (m3/h)/Delta P (Pa)

      150 / 30

    • THE-LA100A

    • Base A (mm)

      305

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      9

    • Airflow (m3/h)/Delta P (Pa)

      300 / 30

    • THE-LA100A

    • Base A (mm)

      457

    • Height B (mm)

      457

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      10,1

    • Airflow (m3/h)/Delta P (Pa)

      335 / 30

    • THE-LA100A

    • Base A (mm)

      457

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      13,5

    • Airflow (m3/h)/Delta P (Pa)

      450 / 30

    • THE-LA100A

    • Base A (mm)

      610

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      18

    • Airflow (m3/h)/Delta P (Pa)

      600 / 30

    • THE-LA100A

    • Base A (mm)

      610

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      22,5

    • Airflow (m3/h)/Delta P (Pa)

      750 / 30

    • THE-LA100A

    • Base A (mm)

      610

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      27

    • Airflow (m3/h)/Delta P (Pa)

      900 / 30

    • THE-LA100A

    • Base A (mm)

      610

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      36

    • Airflow (m3/h)/Delta P (Pa)

      1200 / 30

    • THE-LA100A

    • Base A (mm)

      610

    • Height B (mm)

      1830

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 30

    • THE-LA100A

    • Base A (mm)

      762

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      28,1

    • Airflow (m3/h)/Delta P (Pa)

      935 / 30

    • THE-LA100A

    • Base A (mm)

      915

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      40,5

    • Airflow (m3/h)/Delta P (Pa)

      1350 / 30

    • THE-LA100A

    • Base A (mm)

      915

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 30

  • H13

    • THE-LA130A

    • Base A (mm)

      305

    • Height B (mm)

      305

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      4,5

    • Airflow (m3/h)/Delta P (Pa)

      150 / 75

    • THE-LA130A

    • Base A (mm)

      305

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      9

    • Airflow (m3/h)/Delta P (Pa)

      300 / 75

    • THE-LA130A

    • Base A (mm)

      457

    • Height B (mm)

      457

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      10,1

    • Airflow (m3/h)/Delta P (Pa)

      335 / 75

    • THE-LA130A

    • Base A (mm)

      457

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      13,5

    • Airflow (m3/h)/Delta P (Pa)

      450 / 75

    • THE-LA130A

    • Base A (mm)

      610

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      18

    • Airflow (m3/h)/Delta P (Pa)

      600 / 75

    • THE-LA130A

    • Base A (mm)

      610

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      22,5

    • Airflow (m3/h)/Delta P (Pa)

      750 / 75

    • THE-LA130A

    • Base A (mm)

      610

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      27

    • Airflow (m3/h)/Delta P (Pa)

      900 / 75

    • THE-LA130A

    • Base A (mm)

      610

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      36

    • Airflow (m3/h)/Delta P (Pa)

      1200 / 75

    • THE-LA130A

    • Base A (mm)

      610

    • Height B (mm)

      1830

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 75

    • THE-LA130A

    • Base A (mm)

      762

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      28,1

    • Airflow (m3/h)/Delta P (Pa)

      935 / 75

    • THE-LA130A

    • Base A (mm)

      915

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      40,5

    • Airflow (m3/h)/Delta P (Pa)

      1350 / 75

    • THE-LA130A

    • Base A (mm)

      915

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 75

  • H14

    • THE-LA140A

    • Base A (mm)

      305

    • Height B (mm)

      305

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      4,5

    • Airflow (m3/h)/Delta P (Pa)

      150 / 85

    • THE-LA140A

    • Base A (mm)

      305

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      9

    • Airflow (m3/h)/Delta P (Pa)

      300 / 85

    • THE-LA140A

    • Base A (mm)

      457

    • Height B (mm)

      457

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      10,1

    • Airflow (m3/h)/Delta P (Pa)

      335 / 85

    • THE-LA140A

    • Base A (mm)

      457

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      13,5

    • Airflow (m3/h)/Delta P (Pa)

      450 / 85

    • THE-LA140A

    • Base A (mm)

      610

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      18

    • Airflow (m3/h)/Delta P (Pa)

      600 / 85

    • THE-LA140A

    • Base A (mm)

      610

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      22,5

    • Airflow (m3/h)/Delta P (Pa)

      750 / 85

    • THE-LA140A

    • Base A (mm)

      610

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      27

    • Airflow (m3/h)/Delta P (Pa)

      900 / 85

    • THE-LA140A

    • Base A (mm)

      610

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      36

    • Airflow (m3/h)/Delta P (Pa)

      1200 / 85

    • THE-LA140A

    • Base A (mm)

      610

    • Height B (mm)

      1830

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 85

    • THE-LA140A

    • Base A (mm)

      762

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      28,1

    • Airflow (m3/h)/Delta P (Pa)

      935 / 85

    • THE-LA140A

    • Base A (mm)

      915

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      40,5

    • Airflow (m3/h)/Delta P (Pa)

      1350 / 85

    • THE-LA140A

    • Base A (mm)

      915

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 85

  • U15

    • THE-LA150A

    • Base A (mm)

      305

    • Height B (mm)

      305

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      4,5

    • Airflow (m3/h)/Delta P (Pa)

      150 / 100

    • THE-LA150A

    • Base A (mm)

      305

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      9

    • Airflow (m3/h)/Delta P (Pa)

      300 / 100

    • THE-LA150A

    • Base A (mm)

      457

    • Height B (mm)

      457

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      10,1

    • Airflow (m3/h)/Delta P (Pa)

      335 / 100

    • THE-LA150A

    • Base A (mm)

      457

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      13,5

    • Airflow (m3/h)/Delta P (Pa)

      450 / 100

    • THE-LA150A

    • Base A (mm)

      610

    • Height B (mm)

      610

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      18

    • Airflow (m3/h)/Delta P (Pa)

      600 / 100

    • THE-LA150A

    • Base A (mm)

      610

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      22,5

    • Airflow (m3/h)/Delta P (Pa)

      750 / 100

    • THE-LA150A

    • Base A (mm)

      610

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      27

    • Airflow (m3/h)/Delta P (Pa)

      900 / 100

    • THE-LA150A

    • Base A (mm)

      610

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      36

    • Airflow (m3/h)/Delta P (Pa)

      1200 / 100

    • THE-LA150A

    • Base A (mm)

      610

    • Height B (mm)

      1830

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 100

    • THE-LA150A

    • Base A (mm)

      762

    • Height B (mm)

      762

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      28,1

    • Airflow (m3/h)/Delta P (Pa)

      935 / 100

    • THE-LA150A

    • Base A (mm)

      915

    • Height B (mm)

      915

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      40,5

    • Airflow (m3/h)/Delta P (Pa)

      1350 / 100

    • THE-LA150A

    • Base A (mm)

      915

    • Height B (mm)

      1220

    • Thickness C (mm)

      150

    • Filter pack (mm)

      100

    • Surface (m2)

      54

    • Airflow (m3/h)/Delta P (Pa)

      1800 / 100

* Energy consumption, kWh/year: calculated according to Eurovent Guideline 4/21-2018 | ** Energy class: according to Eurovent RS4/C/001-2019 | *** Energy class: according to Eurovent RS4/C/001-2015
Application areas

An air filtration system adapted to each sector

Illustration

Hospitals

Equip your hospital with air filtration systems adapted to the healthcare environment and your activity.

Discover all sectors
Illustration
TFM® Service

Choose TFM® and benefit from a comprehensive maintenance contract for air filtration.

Need clean air in your workplace? With the TFM® contract, you will benefit from high-performance air filters tailored to your needs, scalable services, and dedicated teams to advise you on the installation of new equipment.

Learn more about our service

Do you have any questions? Our team is here to help.

Our experts are available to answer all your questions.

Contact our expert
Illustration
Expert illustration

Hi, my name is Wannes Van Herck, how can I assist you?

Contact our expert